Charge-balancing topologies guaranteeing zero net DC current deposition into brain tissue.
Charge density per phase is strictly constrained below the Shannon threshold (k = 1.75) to prevent platinum electrode dissolution and tissue excitotoxicity.
After every cathodic stimulation pulse, an active anodic pull-back phase discharges residual capacitive polarization, guaranteeing net DC zero deposition across millions of cycles.